background image

General Software Operation

  Read 

 Loads the data from the connected device (sensor or transmitter) and updates the software’s display. 

 

  NOTE

: This will overwrite any changes made in the 0252 software since the last 

Write

.

  Write

   Applies the data entered in the 0252 software to the connected device. Once you have entered the desired setting 

changes in the software screens, press 

Write

 to load your new settings onto the connected device.

  Save 

 Stores the entire 0252 settings configuration, as currently displayed in the application, to a specified location on your 

computer. (You will be asked to select a file location and provide a file name)

  Load

   Opens a previously saved settings configuration file. See 

Save

 function above.

 

  NOTE:  

The file must be a 0252 settings configuration file. 

 

The software will verify whether the user-selected file is the correct type.

  Default

   Resets all data on all application screens to a factory default condition. A confirmation dialog box will be presented with 

a warning explaining all un-saved configuration information will be erased. 

 

After resetting the software to a factory default condition, click 

Write

 to reset the connected device to a factory default 

condition.

 

  NOTE:  

Performing the above reset to factory default condition will not change the input type on the 

 

9900 Transmitter.

  

NOTE: 

Certain sensors, 2250, 2450, and 258X will request sensor specific information for the default setting 

 

 

   

   such as range or size.

On all screens 

Read, Write, Save, Load, 

and

 Default 

buttons can be found in the lower right corner.

Sensor Operation

Application

 settings:

2250

2350

2450

2750

2751

1

Engineering units

in., ft, cm, m

˚C or ˚F

psi, kpa, bar

pH or (ORP)

pH or (ORP)

Loop

 Output settings:

2

4 mA 

Set Point

Min

0

-10 ˚C (14 ˚F)

0

0 pH 

 

(-1000 mV)

-1 pH

(-1999 mV)

Max

-XU: 

276.8 in., 23.07 ft, 

 

703.1 cm, 7.03 m 
-XL: 

 

1384.0 in, 115.33 ft, 

3515.0 cm, 35.15 m

100 ˚C (212 ˚F)

psi: 10, 50, 250

kpa: (69, 344.7, 

 

1723.7)

bar: (0.7, 3.5, 17.2)

14 pH 

 

(2000 mV)

15 pH 

 

(1999 mV)

3

20 mA 

Set Point

Min

0

-10 ˚C  (14 ˚F)

0

0 pH 

 

(-1000 mV)

-1 pH

(-1999 mV)

Max

-XU: 

 

276.8 in., 23.07 ft, 

 

703.1 cm, 7.03 m 
-XL: 

 

1384.0 in, 115.33 ft, 

3515.0 cm, 35.15 m

100 ˚C (212 ˚F)

psi: 10, 50, 250

kpa: (69, 344.7, 

 

1723.7)

bar: (0.7, 3.5, 17.2)

14 pH 

 

(2000 mV)

15 pH 

 

(1999 mV)

Write

 or 

Save

 settings:

4a

or

4b

Click “Write” to copy these settings to the product.
Click “Save” to save these settings to a local computer file for later use.

5

To use a saved file (from 4b):
1. Click “Load”
2. Navigate to the saved file
3. Select “Open”
4. Click “Write”
5. Click “Read” to confirm

7

GF 0252 Configuration Tool

Summary of Contents for Signet 0252

Page 1: ...ngs to factory default condition Calibrate 258X FlowtraMag Full Bore Magmeter 2610 Dissolved Oxygen Sensor and pH ORP sensors using the 2751 Smart Sensor Electronics Microsoft and Windows are register...

Page 2: ...Blue Indicator DATA COMMUNICATION Input connections 3 terminal connectors max 14 AWG Electrical Communication rate Maximum 19 2 kbs Input power Supplied by USB interface Output power 5 VDC 5 Power con...

Page 3: ...the drop down menu Supported languages English French German Spanish Italian Portuguese Chinese 5 Click Save to set language preference Caution Managed systems and network systems may have security m...

Page 4: ...ndows PC Black Red White 1 2 3 4 White Red Black S USB GND Data 5V out USB GND Data 5V out Windows PC Windows PC Black Red White 1 2 3 4 White Red Black S USB GND Data 5V out USB GND Data 5V out Windo...

Page 5: ...ire 1 Additional wire required Brown Black White Blue White Blue Wire Nut To VDC To GND Wire 3 Wire 2 USB GND Data 5V out Windows PC Brown Black White Blue USB GND Data 5V out Windows PC Isolated 24 V...

Page 6: ...ower Supply 12 to 32 VDC Brown Blue Green No connection CAUTION Do not connect power directly to the 0252 Configuration Tool Improper power supply wiring can cause damage to the 0252 hardware and comp...

Page 7: ...reset the connected device to a factory default condition NOTE Performing the above reset to factory default condition will not change the input type on the 9900 Transmitter NOTE Certain sensors 2250...

Page 8: ...ut Off of full scale 0 0 to 20 Full Scale Loop Output settings 11 4 mA Set Point Flow Units Time Base 0 00000 0 00000 999999 00000 12 20 mA Set Point Flow Units Time Base 5 00000 Write or Save setting...

Page 9: ...allow the Magmeter to override AVERAGING and jump to a new flow rate immediately 2551 maximum range is 10 m s See Magmeter manual for an explanation of Averaging and Sensitivity www gfps com Noise Rej...

Page 10: ...Select Output modes 4 to 20 mA as either Active or Passive S L or Frequency output Note Setting the 258X to frequency will prevent the use of the 0252 tool You will need to change the 258X back to S...

Page 11: ...258X Operation 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 11 GF 0252 Configuration Tool...

Page 12: ...low Rate settings 6 Wait for flow reading to stabilize then enter correct flow and Click Set Flow 7 Cancel Flow Rate Calibration and return to Calibration screen Volumetric settings 8 Start Volumetric...

Page 13: ...258X Calibration 6 7 8 9 10 11 13 GF 0252 Configuration Tool...

Page 14: ...Water Salinity Value for Salt Compensation 0 Fresh Water Loop Output settings 5 Select Measurement Dissolved Oxygen or Temperature for 4 to 20 mA Output 6 Set 4 mA Set Point 7 Set 20 mA Set Point Inf...

Page 15: ...ated Water click Start to begin the Calibration 4 Once probe reading is stable click Continue Second Point settings 5 With probe in 0 Saturated Water click Start to begin the second point Calibration...

Page 16: ...ss Impedance Set Point for Alarm pH 10 Glass Impedance Update Timer pH Information settings 11 Click Get Data to Update Sensor Data 12 Electrode Serial Number pH or ORP 13 Electrode Part Number pH or...

Page 17: ...2751 Operation 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 17 GF 0252 Configuration Tool...

Page 18: ...Click Next for two point calibration 9 Click End to complete one point calibration 10 To cancel calibration click Cancel and return to Calibration Screen Temperature Calibration settings 11 Start Cali...

Page 19: ...2751 Calibration 5 6 7 8 9 10 11 12 19 GF 0252 Configuration Tool...

Page 20: ...2751 Calibration 13 14 15 16 17 20 GF 0252 Configuration Tool...

Page 21: ...on settings 3 Select the Input Calibration Loop Relay or Option tabs to choose the desired menu for the selected sensor 4 Refer to the 9900 manual for details pertaining to specific settings for each...

Page 22: ...9900 Operation 2 3 5a 5b General Software Operation pg 7 22 GF 0252 Configuration Tool...

Page 23: ...the maximum interval is 86400 seconds 24 hours Example If the Log Interval is set to 60 seconds the 0252 will record the temperature once every minute 4 The 0252 saves data files in csv format The ma...

Page 24: ...ce visit our website www gfps com Or call in the U S 800 854 4090 For the most up to date information please refer to our website at www gfps com 3 0252 090 Rev 3 01 22 English Georg Fischer Signet LL...

Reviews: