4 Assembly and Installation
Description to tubing connections (Fig. 5.4.a)
1
PVC-pipe 6 x 1 mm, 400 mm long
2
PVC-pipe 6 x 1 mm, 400 mm long
3
PVC-tubing 4/7, 2 m, inlet H
2
O steam generator
4
PVC-tubing 4/7, 2 m, inlet H
2
O sample
5
PVC-tubing 4/7, 2 m, inlet H
3
BO
3
6
PVC-tubing 8/12, 2 m inlet NaOH
7
Cooling water inlet tubing 10/17, 2 m lang
8
PVC-outlet tubing 8/12, 2 m, for cooling water (place into a drain)
9*
PVC-outlet tubing 8/12, 2 m, for receiver (place into a drain)
*only for models with external titrator
10
Verprene-tubing white 8/12, 2 m for sample waste
11
PVC-pipe 10 x 1 mm, 400 mm long
12
PVC-pipe 6 x 1 mm, 400 mm long
13
Storage tank 20 l, for sample waste
14
Storage tank 20 l, for NaOH (Reagent)
15
Storage tank 20 l, for H
3
BO
3
16
Storage tank 20 l, for H
2
O
Connection of set of storage tank KAN 40 / KAN 50
A
4 x connection cable from level detection tank to distribution box
B
Distribution box with 4 plugs (KAN 50 with 5 plugs)
C
Connection cable from distribution box to socket level detection Vapodest
D
Level sensor with plug for sample waste tank, complete
E
Level sensor with plug for chemicals tank, complete
Fill tanks with chemicals:
- H
2
0 distilled or demineralizedt
- e.g. H
3
BO
3
2 - 4 %
- e.g. NaOH 32%
- Titration acid
4.8. Mains connection Vapodest
Please check the nominal voltage on the product information plate before con-
necting to the mains. Should the nominal voltage deviate for more than +/- 10%,
then you must contact your service engineer.
- Connect the mains cable with the rear appliance plug and the shockproof
socket
- Check if excess current release at the rear (chapter 4.6., pos. 7) of the equip-
ment is in position „1“. This switch turns the equipment off when it takes too
much current.