35
Vapodest 45s
M01 Adding H3BO3
80 ml
8 Execution of method
M01 Ammonium
Selection
turn
press
6. Program number is flashing.
Select the method program as desired (M00 to
M19).
7.
Continue = the program starts immediately
Stop = back to main menu
Current selection is flashing
In the display, the individual program steps are
shown.
The program can be interrupted at any time during
the distillation in progress. Please see chapter
"8.4. Interrupt distillation"
.
8
. At the end of the program, take out the digestion
tube. In order to do so, push the quick clamping
device downwards.
Careful ! At the end of the distillation, the digestion
tube is very hot. Thus make sure to wear protective
gloves and protective goggles!!
9.
Would you like to continue with the program and
run additional analysis, then please insert a new
digestion tube and press the push button. The
receiver will be filled with boric acid again.
Should you not wish any additional analysis, then
turn the push button in any direction and select
"Exit".
press
Method
Execution
press
M01 Ammonium
Stop Continue
M01 Adding H2O
100ml
M01 Adding NaOH
080ml
M01 Reaction
00min 00s
M01 Distillation
03min 30s
M01 Suction
P: 25 s
M01 Ammonium
Finished
press
M01 Ammonium
Stop Continue
M01 Ammonium
Exit
turn