10 Rework instruction
10.1. Upgrade the Vap 45s to be operated with external titrator
In case you have ordered the Vapodest 45 without titrator and now, wish to up-
grade your unit to be used with an external titrator, we kindly ask you to follow the
instructions enclosed:
- Take off platform pos 2. and PVC tube holder
pos. 1
- Slide the fitting of the receiver pos. 3 on the fitting
and attach it with the screw enclosed (M3 x 20).
- Attach receiver glass pos. 8 under the receiver
table. Move the receiver glass for about 90°, in
order to attach it.
- Silicone tubing for the addition of boric acid pos.
7 and Slide silicone tubing of the distillate outlet
tubing pos. 4 through the opening into the titration
cell (shorten tubing so that they will end shortly
above the glass bottom).
- Remove protective cap from the connection pos.
6. Put Verprene tubing for the suction on the
connection and join with the respective Teflon
connection.
- Establish plug-in connection:
Connect 5-pos. Drehverschlussstecker pos. 5
- Set up titrator and connect:
See chapter "4.9. Connect titrator and electrode"
- Connect electrode.
See chapter "4.9. Connect electrode and titrator“
- Change system settings
To make sure that the Vapodest recognizes that
an external titrator has been connected, you must
make a selection in the menu "Titration".
See chapter "6.5. Titration