33
Vapodest 45s
M01 Ammonium
Time: XXmin XXs
press
M01 Suction time
Receiver 25s
turn
press
press
9. Suction time receiver
(only in combination with external titrator)
The duration of the suction is calculated so that
the receiving vessel is completely empty. Should
there be any remaining single drops, they won’t
have any influence on the following analyses.
- Setting range 0 - 99 s
10. Adition H
3
BO
3
The diaphragm of the electrode must be completely
immersed in the receiving solution, so that a stable
pH- measurement is ensured.
-
Setting range 0 - 200 ml
(the specification of the dosing steps depends on
the calibration of the pumps)
11
. Total time
Display shows the total time for this program.
M01 Adding H3BO3
80 ml
turn
press
press
Method
Edit
7 Create a method (Programming)