The functions of the distillation system are monitored continuously. As soon as
an error occurs, it is shown in the Vapodest-display, After solving the problem,
the analysis can be continued in most cases. The user might have to confirm a
corresponding message in the display.
9.1. Error messages
9 Trouble shooting
Water pressure
low
Error message
Reason for error
Actions to be taken
Cooling water pressure
< 0,5 bar.
Check cooling water inlet. Make sure that the faucet
is turned on.
Tube or door
Error
No sample tube or door
open
Insert digestion tube and / or close protection door..
Distillations can only be done if the digestion tube is
inserted correctly and the protection door is closed.
Wait for steam
Steam generator has not
yet reached its required
pressure
Message disappears, as soon as the steam generator
has reached the correct pressure.
Method
Not defined
The selected program has
not yet been defined
Define method program or select a different method
Tank level
Error
One or more storage
tanks are empty or samp-
le waste tank is full.
- Check content of storage tanks and if necessary,
fill it up.
- Also, check contacts for level sensor
- The distillation in process will be completed.
Titration time
>15min
Titrator not
ready
Receiver
Overfilled
- Check whether titrator
is connected correct-
ly
- Cable to titrator might
be broken
- Check cable connecting titrator with Vapodest
- Use new cable for titrator
- Sucction for receiver
manually
Press combination of keys.
+
Electrode is too slow or
broken
Calibrate electrode or replace.