PARAMETER SETTING PRIOR TO DELIVERY
The damping value (time constant), function of zero/span adjust screw, output current mode,
indicator scale, cut point, mode below cut point and burnout, have been set prior to delivery as
shown in the following.
Each parameter is changed by using HHC or the 3 push button of the LCD display.
N°
Item
Contents of parameter
1
Damping value
(time constant)
0,06 sec
2
External adjustment function
of the transmitter
Adjustable (ENABLE)
3
Current output mode
Linear (Could be set in factory
when ordering (note 2))
Digital indicator scale
(9th digit of code symbol)
Could be set in factory
when ordering
4
Cut point
(square root extraction mode setting)
7.07%
5
Mode below cut point
(square root setting)
Linear
6
Burnout
Hold (note 3)
7
Polygonal line correction
Not corrected (INVALID)
8
Saturation current
Normal specification (NORMAL)
9
Protective function of set value
(write protect)
Cancelled (OFF)
Note 1) Use the HHC or the local configurator unit with LCD display to change the setting of all the items
except “7: Polygonal line correction,” which can be changed only by the HHC.
Note 2) In both the differential pressure transmitter (Type: FKC) and remote seal type (Type: FKD), the output
current mode is set in linear unless it is designated.
Note 3) “Hold” is selected for “Burnout” unless otherwise specified by the order.