Voluson Format(*.4dv)
•
V730 format (V730.mdb)
Using Import
1.
Press
Import
to start importing data.
2.
Select
•
the data type
•
the location of the file you want to import data from
•
import preview from USB
3.
Confirm by pressing
Open
.
Import preview from USB
It is possible to review and/or reload data directly from the selected location. In reload state
you can
Send
and
but not
Save
. When
Save
is selected following message pops up:
"SAVE not possible! Exam must be imported first".
After closing this window with "OK" another dialog appears. It tells the user to press
Import
again should he want to copy the data to the hard drive permanently (restore).
Possible to import:
•
selected patients
•
selected exams
•
selected image and volume data
All data contained in the 4DV file or the V730.mdb file is displayed in the Archive window
under the source category “4DV”. The user can now select the patients and exams that he
wants to import.If an encrypted 4DV file is selected, a password prompt appears. The entering
of a wrong password causes an error message.
Selecting the
Import
again and copying the data to the hard drive. If files are missing/corrupted
and cannot be imported/restored, a warning appears.
Importing from a DICOM Server (Query Retrieve)
Exams, patients and images from a DICOM Server can be imported after
Query/Retrieve was performed. It is only possible to import complete exams
or patients, not single selected images. If no images are available, the
Import
button is disabled.
9.2.5 Delete
Deleting patients, images or exams
1.
Select the data (patient(s), exam(s) and/or image(s)) to delete.
2.
Press
Delete
on the screen.
3.
A dialog appears which asks for confirmation to delete the selected items.
Controls
Delete Images only
Selected image(s) will be deleted.
Delete All Data
All selected data will be deleted.
No
Cancels deleting data.
Archive