Section 3-5 - Connection of Auxiliary Devices
Section 3-5 Connection of Auxiliary Devices
(cont’d)
1.) Carefully remove plastic caps using a knife and loosen screws and washers (see:
Figure 3-6
).
2.) When the cover is loose on top, pull the rear cover out and move upwards.
3.) Disconnect the GND ground-cable from the back of the rear cover plate.
4.) Connect all Peripherals according to connection scheme; see:
Figure 3-5
.
5.) Connect the GND ground-cable at the rear cover plate.
6.) Mount rear cover plate, reattach screw and washers and plug-in the caps.
Figure 3-6 Remove rear cover plate
Figure 3-7 GND ground-cable
plastic caps; screw and washers
GND ground cable