4 Operation
4.4 Performing a scanning
42
Typhoon
Operating Instructions
28-9623-56
AA
•
For a transparent sample (such as a polyacrylamide gel), place the
sample either face up or face down.
•
If the sample is physically uneven on one side (such as an agarose gel),
place the smooth side down. This allows the sample to lie flat.
b) Figure below shows a top view of instrument with instrument front (A) and
glass platen (B).
The sample (C) should be positioned in the lower left corner (A1) of the scan
area. Each lettered or numbered grid segment is marked by two dots.
c)
Align the lanes and rows of the sample with the edges of the grid on the glass
platen so that the scanned image will be straight.
d) Make a note of the coordinates of the upper left and lower right corners of the
area you want to scan (see step 4b in this instructions for an example). This
information will be used when setting the scanning area.
5
Close the sample lid and press it down until the latch clicks. The screen or sample
is now ready for scanning.
A
1
B
C
D
E
F
G
H
I
J
K
L
M
N
O
P
Q
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
P
Q
C
A
A1
B
Summary of Contents for Typhoon 9200
Page 2: ...GE Healthcare Typhoon Operating Instructions Original instructions...
Page 3: ......
Page 13: ...1 Introduction 1 4 Control software 12 Typhoon Operating Instructions 28 9623 56 AA...
Page 27: ...3 Installation 3 5 Spare parts and accessories 26 Typhoon Operating Instructions 28 9623 56 AA...
Page 75: ...Appendix A Accessories A 4 Beamsplitters 74 Typhoon Operating Instructions 28 9623 56 AA...
Page 76: ......