4 Operation
4.4 Performing a scanning
48
Typhoon
Operating Instructions
28-9623-56
AA
a) Click the
Orientation
button
b) Click the button that represents how the sample is oriented.
c)
During the scan, the Typhoon Scanner Control software maps the pixels to
display the image of your sample face up and top-end up.
8
Optional for fluorescent samples:
Select Press Sample
parameter (
, H).
Tick the
Press Sample
check box if a sample that does not lie flat against the glass
platen of the instrument, such as a dry membrane or filter paper, is to be scanned.
The inner lid of the
Typhoon
instrument will go down to press the sample flat.
Note:
Do not select
Press Sample
if you are scanning wet membranes, wet gels, or
soft samples. Pressing wet or soft samples can damage the instrument. In
addition, pressing a soft sample can distort the scanned image.
Note:
In the
Storage phosphor scan
acquisition mode, the
Press Sample
parameter will always be used.
9
Select pixel size from the
Pixel size
list (
, I).
Pixel size
(µm)
Recommendations
1 000
For a very low resolution scan of the screen.
Produces 10 data points per centimeter and 20 data lines per
grid square.
Note:
The 1 000-µm pixel size produces very rough data that is
not recommended for quantitative analysis.
500
For a low resolution scan of the screen.
Produces 20 data points per centimeter and 40 data lines per
grid square.
200
For most standard electrophoresis samples.
Produces 50 data points per centimeter and 100 data lines per
grid square.
100
For samples requiring high resolution (such as DNA sequences).
Produces 100 data points per centimeter and 200 data lines per
grid square.
Summary of Contents for Typhoon 9200
Page 2: ...GE Healthcare Typhoon Operating Instructions Original instructions...
Page 3: ......
Page 13: ...1 Introduction 1 4 Control software 12 Typhoon Operating Instructions 28 9623 56 AA...
Page 27: ...3 Installation 3 5 Spare parts and accessories 26 Typhoon Operating Instructions 28 9623 56 AA...
Page 75: ...Appendix A Accessories A 4 Beamsplitters 74 Typhoon Operating Instructions 28 9623 56 AA...
Page 76: ......