Servicing the PT878GC
9-45
April 2009
Updating Software Via IrOBEX (cont.)
5.
The meter now asks if you wish to reload the flash memory
via the infrared communications port.
•
Press
[F1]
(Yes) to begin reloading via OBEX, or
•
Press
[F2]
(No) to stop the reloading procedure.
Be sure the infrared sensor on the meter has a clear path to the
infrared adapter on the PC. Now return to the PC.
6.
You now have two options:
•
In Windows Explorer, browse to the file you wish to send.
Select and right-click on it. Select
Send To
Infrared
Recipient
or
Nearby Computer
(or
QuickBeam
if
QuickBeam has been installed)
.
•
Open the
Infrared Transfer Application
(
My
Computer\Infrared Recipient
), click on
Send Files,
and
select the files you wish to send.
In either case, a window displays the status of the file upload.