Servicing the PT878GC
Installing New Software (cont.)
1.
As shown in Figure 9-30 on the previous page, the Flash
Update option erases all instrument coding.
•
Press
[F2]
(Cancel) to return to Operate Mode without
erasing the program.
•
Press
[F3]
(OK) to confirm that you wish to erase the
program.
2.
The meter asks for confirmation. Repeat the options shown in
Step 1 above. The screen on both the PC and the PT878GC
now appears similar to Figure 9-31 below.
Figure 9-31: Flash Reloading Window
3.
The meter now asks if you wish to reload the flash memory
via the infrared communications port.
•
Press
[F3]
(Yes) (or C on the PC keyboard) to begin
reloading via IrCOMM, or
•
Press
[F2]
(No) (or ESC on the keyboard) to stop the
reloading procedure.
4.
Be sure the infrared sensor on the meter has a clear path to the
infrared adapter on the PC, and return to the PC
Hyperterminal program.
GE Sensing PCI Loader v3.0 2/26/04
[HW Rev3+]
Reload Flash via Infrared (Yes/No)?
Backup Battery: OK
Firmware CRC: Valid
IrOBEX No IrCOMM