Appendix A–Technical Specifications
Invasive blood pressure
Number of channels:
Up to 4 (with appropriate cables)
Transducer sites, site name
and displayed values:
arterial (ART): systolic, diastolic, mean and rate
femoral (FEM): systolic, diastolic, mean and rate
pulmonary artery (PA): systolic, diastolic and mean
central venous pressure (CVP): mean
left atrial (LA): mean
right atrial (RA): mean
intracranial pressure (ICP): mean
umbilical artery (UAC): systolic, diastolic, mean and rate
umbilical vein (UVC): mean
special pressure (SP): mean
Transducer requirements
Transducers meeting the following specifications can be
used, however, have not been evaluated by UL.
NOTE
Note: UL Classification covers the following invasive
blood pressure transducers:
Utah, Deltran IV CPT-400
Edwards, PX600
Bekton Dikinson, P23XL-1
Abbot, 42582-08
Accuracy:
±
2% or
±
1 mmHg, whichever is greater (exclusive of
transducer)
±
2% or
±
3 mmHg, whichever is greater (exclusive of Patient
Data Module)
Over pressure rating:
500 kPa (4000 mmHg) above atmospheric pressure (all
except catheter tip transducers)
Under pressure rating:
50 kPa (400 mmHg) below atmospheric pressure (all except
catheter tip transducers)
Excitation voltage:
Transducer output:
±2.5 V dc
±
0.1%
50
μ
V/V/cm Hg
Input specifications
Range:
Offset:
-25 mmHg to 349 mmHg
±150 mmHg
Summary of Contents for PDM
Page 7: ...2030047 001A Patient Data Module 1 1 1 Introduction ...
Page 13: ...2030047 001A Patient Data Module 2 1 2 Equipment Overview ...
Page 21: ...2030047 001A Patient Data Module 3 1 3 Installation ...
Page 26: ...3 6 Patient Data Module 2030047 001A Installation ...
Page 27: ...2030047 001A Patient Data Module 4 1 4 Service Tool and Configuration ...
Page 38: ...4 12 Patient Data Module 2030047 001A Service Tool and Configuration ...
Page 39: ...2030047 001A Patient Data Module 5 1 5 Maintenance ...
Page 48: ...5 10 Patient Data Module 2030047 001A Maintenance ...
Page 49: ...2030047 001A Patient Data Module 6 1 6 Troubleshooting ...
Page 54: ...6 6 Patient Data Module 2030047 001A Troubleshooting ...
Page 55: ...2030047 001A Patient Data Module 7 1 7 Field Replaceable Units ...
Page 69: ...2030047 001A Patient Data Module A 1 A Appendix A Technical Specifications ...
Page 80: ...A 12 Patient Data Module 2030047 001A Appendix A Technical Specifications ...
Page 81: ...2030047 001A Patient Data Module B 1 B Appendix B Electromagnetic Compatibility ...
Page 90: ...B 10 Patient Data Module 2030047 001A Appendix B Electromagnetic Compatibility ...
Page 91: ......