A-2
Patient Data Module
2030047-001A
Appendix A–Technical Specifications
Technical Specifications
Physical
Environmental
Battery
Height:
7.0 cm (2.75 in)
Width:
14.6 cm (5.75 in)
Depth:
21.6 cm (8.5 in)
Weight:
1.1 kg (2.5 lbs.) without optional battery
1.3 kg (2.8 lbs.) with optional battery
Power requirements
Input voltage:
Input current:
8 to 17 Vdc
0.3A nominal at 16.75 V
Power consumption:
4.5 Watts, nominal
Cooling:
Natural convection
Operating heat dissipation:
10.2 Btu/Hr
NOTE
System may not meet its performance specifications if stored or used outside the
manufacturer’s specified temperature and humidity ranges.
Operating temperature and
humidity:
10º C to 35º C (50º F to 95º F)
15% to 95% RH non-condensing
Storage temperature and
humidity:
-40º C to 60º C (-40º F to 140º F)
15% to 95% RH non-condensing
Type:
Exchangeable lithium ion
Quantity:
One
Voltage:
11.1 V nominal
Capacity:
1.8 Amp hour nominal
Run time:
Approximately 3.5 hours with new fully charged battery
Charge time:
Approximately 2 hours
Battery life:
500 cycles to 50% capacity
Summary of Contents for PDM
Page 7: ...2030047 001A Patient Data Module 1 1 1 Introduction ...
Page 13: ...2030047 001A Patient Data Module 2 1 2 Equipment Overview ...
Page 21: ...2030047 001A Patient Data Module 3 1 3 Installation ...
Page 26: ...3 6 Patient Data Module 2030047 001A Installation ...
Page 27: ...2030047 001A Patient Data Module 4 1 4 Service Tool and Configuration ...
Page 38: ...4 12 Patient Data Module 2030047 001A Service Tool and Configuration ...
Page 39: ...2030047 001A Patient Data Module 5 1 5 Maintenance ...
Page 48: ...5 10 Patient Data Module 2030047 001A Maintenance ...
Page 49: ...2030047 001A Patient Data Module 6 1 6 Troubleshooting ...
Page 54: ...6 6 Patient Data Module 2030047 001A Troubleshooting ...
Page 55: ...2030047 001A Patient Data Module 7 1 7 Field Replaceable Units ...
Page 69: ...2030047 001A Patient Data Module A 1 A Appendix A Technical Specifications ...
Page 80: ...A 12 Patient Data Module 2030047 001A Appendix A Technical Specifications ...
Page 81: ...2030047 001A Patient Data Module B 1 B Appendix B Electromagnetic Compatibility ...
Page 90: ...B 10 Patient Data Module 2030047 001A Appendix B Electromagnetic Compatibility ...
Page 91: ......