7-12
Patient Data Module
2030047-001A
Field Replaceable Units
7. Licensing information must be re-entered. Contact GE technical support and
provide the Patient Data Module serial number and MAC address to proceed. See
the “How to Reach Us...” page included with this manual for contact information.
Go to
for instructions on entering the new activation code.
8. Manufacturer’s serial number must be re-entered. Go to
for instructions on entering the manufacturer’s serial number.
9. Go to the Solar 8000M/i patient monitor service manual and complete checkout
procedures.
NBP hose, coupling and elbow
1. Remove the top housing and main board per above steps.
2. Disconnect the front panel connector.
477A
3. Disconnect the main board flex pump connector.
4. Hold flex out of the way and disconnect the NBP hose from the coupling.
5. Remove hose from the manifold.
NOTE
When reassembling, slide hose all the way onto the manifold fitting so that
the hose lays flat.
6. Go to the Solar 8000M/i patient monitor service manual and complete checkout
procedures.
Front panel
connector
Main board flex pump
connector
NBP hose
and coupling
NBP hose at
manifold
Summary of Contents for PDM
Page 7: ...2030047 001A Patient Data Module 1 1 1 Introduction ...
Page 13: ...2030047 001A Patient Data Module 2 1 2 Equipment Overview ...
Page 21: ...2030047 001A Patient Data Module 3 1 3 Installation ...
Page 26: ...3 6 Patient Data Module 2030047 001A Installation ...
Page 27: ...2030047 001A Patient Data Module 4 1 4 Service Tool and Configuration ...
Page 38: ...4 12 Patient Data Module 2030047 001A Service Tool and Configuration ...
Page 39: ...2030047 001A Patient Data Module 5 1 5 Maintenance ...
Page 48: ...5 10 Patient Data Module 2030047 001A Maintenance ...
Page 49: ...2030047 001A Patient Data Module 6 1 6 Troubleshooting ...
Page 54: ...6 6 Patient Data Module 2030047 001A Troubleshooting ...
Page 55: ...2030047 001A Patient Data Module 7 1 7 Field Replaceable Units ...
Page 69: ...2030047 001A Patient Data Module A 1 A Appendix A Technical Specifications ...
Page 80: ...A 12 Patient Data Module 2030047 001A Appendix A Technical Specifications ...
Page 81: ...2030047 001A Patient Data Module B 1 B Appendix B Electromagnetic Compatibility ...
Page 90: ...B 10 Patient Data Module 2030047 001A Appendix B Electromagnetic Compatibility ...
Page 91: ......