5-2
Patient Data Module
2030047-001A
Maintenance
Maintenance schedule
Manufacturer recommendations
To help ensure the equipment remains in proper operational and functional order,
adhere to a good maintenance schedule. The manufacturer recommends that the
following be performed by service personnel upon receipt of the equipment, every 12
months thereafter, and every time the unit is serviced:
Visual inspection
Cleaning
Battery maintenance
Electrical safety tests (See the Solar 8000M/i patient monitor v5 service manual,
Maintenance and checkout chapter for system tests.)
Checkout procedures (See the Solar 8000M/i patient monitor v5 service manual,
Maintenance and checkout chapter for system tests.)
Manufacturer responsibility
WARNING
Failure on the part of all responsible individuals, hospitals or
institutions, employing the use of this device, to implement the
recommended maintenance schedule may cause equipment failure
and possible health hazards. The manufacturer does not, in any
manner, assume the responsibility for performing the recommended
maintenance schedule, unless an Equipment Maintenance
Agreement exists. The sole responsibility rests with the individuals,
hospitals, or institutions utilizing the device.
Inspection
Remove power and all cables before inspecting or cleaning the equipment.
Inspect the equipment and its components carefully prior to installation, once every
12 months thereafter and each time the equipment is serviced. Do not use the
equipment if damage is determined. Refer damaged equipment to qualified service
personnel.
Inspect the case for cracks or other physical damage.
Inspect cables for fraying or other damage.
Inspect all plugs and connectors for bent pins or other damage.
Check for loose or missing screws on the mounting hardware.
NOTE
Damaged cables or equipment should be replaced by service personnel.
Summary of Contents for PDM
Page 7: ...2030047 001A Patient Data Module 1 1 1 Introduction ...
Page 13: ...2030047 001A Patient Data Module 2 1 2 Equipment Overview ...
Page 21: ...2030047 001A Patient Data Module 3 1 3 Installation ...
Page 26: ...3 6 Patient Data Module 2030047 001A Installation ...
Page 27: ...2030047 001A Patient Data Module 4 1 4 Service Tool and Configuration ...
Page 38: ...4 12 Patient Data Module 2030047 001A Service Tool and Configuration ...
Page 39: ...2030047 001A Patient Data Module 5 1 5 Maintenance ...
Page 48: ...5 10 Patient Data Module 2030047 001A Maintenance ...
Page 49: ...2030047 001A Patient Data Module 6 1 6 Troubleshooting ...
Page 54: ...6 6 Patient Data Module 2030047 001A Troubleshooting ...
Page 55: ...2030047 001A Patient Data Module 7 1 7 Field Replaceable Units ...
Page 69: ...2030047 001A Patient Data Module A 1 A Appendix A Technical Specifications ...
Page 80: ...A 12 Patient Data Module 2030047 001A Appendix A Technical Specifications ...
Page 81: ...2030047 001A Patient Data Module B 1 B Appendix B Electromagnetic Compatibility ...
Page 90: ...B 10 Patient Data Module 2030047 001A Appendix B Electromagnetic Compatibility ...
Page 91: ......