Troubleshooting
2030047-001A
Patient Data Module
6-3
Problems and solutions
Problem
Solution
Battery is not charging
Check the Patient Data Module’s battery interface as follows.
1.
Check for damage to the battery tray. If damaged, replace the battery tray. See chapter 7,
for ordering parts and replacement procedures.
2.
Check for damage to the battery connector inside the battery compartment. If damaged, contact GE
technical support to service the Patient Data Module. See the “How to Reach Us...” page included with
this manual for contact information.
3.
Check that the battery is charging by inserting a known good battery in the Patient Data Module. If not
charging, contact GE technical support to service the Patient Data Module. See the “How to Reach
Us...” page included with this manual for contact information.
Check the Patient Data Module’s battery as follows.
1.
Check that the battery is fully charged by pressing the
TEST
button on the battery. If the LED is not at
100%, go to
for further charging and conditioning procedures.
2.
Check for damage to the battery connector and external surfaces. If damaged, replace the battery.
3.
Check that the battery is charging with the GE approved battery charger. See
for procedures. If not charging, replace the battery.
External surface is
damaged
Check the Patient Data Module’s top housing as follows.
1.
Check for cracks or other damage to the top housing. If damaged, replace the top housing. See chapter
7,
for ordering parts and replacement procedures.
2.
Check for cracks or other damage to the bottom housing. If damaged, contact GE technical support to
service the Patient Data Module. See the “How to Reach Us...” page included with this manual for
contact information.
NBP is not functioning
Check the NBP calibration date and time as follows.
1.
Use the GEHC MS service tool (See chapter 4,
Service Tool and Configuration
) and go to the
Device
Information
module.
2.
If NBP is not calibrated, perform the NBP calibration according to the procedures in the Solar 8000M/i
service manual, Maintenance and checkout chapter.
Check the NBP performance as follows.
1.
Go to the Solar 8000M/i patient monitor service manual Maintenance and checkout chapter and
perform the NBP calibration check.
2.
If NBP pressure is leaking (not maintaining
CUFF
pressure for at least one minute), check for leakage
on the NBP cuff. (Listen for hissing as air escapes.)
3.
If the cuff is not leaking, use the GEHC MS service tool (See chapter 4,
Service Tool and Configuration
and go to
Diagnostics
>
Service Metrics
(See
). Record the NBP statistics
information for trending.
4.
Replace the NBP assembly with manifold. See chapter 7,
and replacement procedures.
Summary of Contents for PDM
Page 7: ...2030047 001A Patient Data Module 1 1 1 Introduction ...
Page 13: ...2030047 001A Patient Data Module 2 1 2 Equipment Overview ...
Page 21: ...2030047 001A Patient Data Module 3 1 3 Installation ...
Page 26: ...3 6 Patient Data Module 2030047 001A Installation ...
Page 27: ...2030047 001A Patient Data Module 4 1 4 Service Tool and Configuration ...
Page 38: ...4 12 Patient Data Module 2030047 001A Service Tool and Configuration ...
Page 39: ...2030047 001A Patient Data Module 5 1 5 Maintenance ...
Page 48: ...5 10 Patient Data Module 2030047 001A Maintenance ...
Page 49: ...2030047 001A Patient Data Module 6 1 6 Troubleshooting ...
Page 54: ...6 6 Patient Data Module 2030047 001A Troubleshooting ...
Page 55: ...2030047 001A Patient Data Module 7 1 7 Field Replaceable Units ...
Page 69: ...2030047 001A Patient Data Module A 1 A Appendix A Technical Specifications ...
Page 80: ...A 12 Patient Data Module 2030047 001A Appendix A Technical Specifications ...
Page 81: ...2030047 001A Patient Data Module B 1 B Appendix B Electromagnetic Compatibility ...
Page 90: ...B 10 Patient Data Module 2030047 001A Appendix B Electromagnetic Compatibility ...
Page 91: ......