A-4
Patient Data Module
2030047-001A
Appendix A–Technical Specifications
Heart rate
Heart rate averaging: 8 / 4 beats
Display update interval: 2 seconds
Heart rate accuracy:
±
1
%
or
±
1 beat/minute, whichever is
greater
Response time: < 6 seconds
Limit alarm delay: <10 seconds after limit alarm condition
exceeded
Heart rate alarm range: 0 to 300 beats/minute, high limit >
low limit
Arrhythmia analysis: 1 to 100 PVCs/minute
Method: QRS morphology classification and timing based on
single- or multiple-lead analysis
Arrhythmia calls: Full, lethal only, or no arrhythmia
PVC alarm limit range: 1 to 100 PVCs/minute
ST Segment analysis
Measurement description:
ST segment deviation is measured and displayed for all
acquired leads
ST display:
Lead label, ST deviation, current complex superimposed
over a reference complex, J-point indicator and 15-minute
mini-trends are shown for all acquired leads
Measurement point:
Measured at user-selectable measurement points (0, 30, 40,
50, 60, and 80 ms) following the J point
Measurement range:
Display resolution:
ST measurement
averaging:
-12.0 mm to + 12.0 mm
0.1 mm
16 beats
ST alarm limits:
± 12 mm, high limit > low limit, for any event within a lead
group (inferior, lateral or anterior) that exceeds the alarm
limit for that group
Pace detection/rejection
Input voltage range:
Input pulse width:
Rise time:
Over/under shoot:
Mode:
Standard leads available:
±2 mV to ±700 mV
0.1 ms to 2 ms
10
μ
s to 100
μ
s
2 mV (maximum)
Pacemaker artifact rejection ‘On’ or ‘Off’.
I, II, RL, LL
Summary of Contents for PDM
Page 7: ...2030047 001A Patient Data Module 1 1 1 Introduction ...
Page 13: ...2030047 001A Patient Data Module 2 1 2 Equipment Overview ...
Page 21: ...2030047 001A Patient Data Module 3 1 3 Installation ...
Page 26: ...3 6 Patient Data Module 2030047 001A Installation ...
Page 27: ...2030047 001A Patient Data Module 4 1 4 Service Tool and Configuration ...
Page 38: ...4 12 Patient Data Module 2030047 001A Service Tool and Configuration ...
Page 39: ...2030047 001A Patient Data Module 5 1 5 Maintenance ...
Page 48: ...5 10 Patient Data Module 2030047 001A Maintenance ...
Page 49: ...2030047 001A Patient Data Module 6 1 6 Troubleshooting ...
Page 54: ...6 6 Patient Data Module 2030047 001A Troubleshooting ...
Page 55: ...2030047 001A Patient Data Module 7 1 7 Field Replaceable Units ...
Page 69: ...2030047 001A Patient Data Module A 1 A Appendix A Technical Specifications ...
Page 80: ...A 12 Patient Data Module 2030047 001A Appendix A Technical Specifications ...
Page 81: ...2030047 001A Patient Data Module B 1 B Appendix B Electromagnetic Compatibility ...
Page 90: ...B 10 Patient Data Module 2030047 001A Appendix B Electromagnetic Compatibility ...
Page 91: ......