5-6
Patient Data Module
2030047-001A
Maintenance
4. Wipe off cleaning solution with a clean, moist swab.
5. Dry thoroughly with a clean, dry, swab and let air dry for at least 30 minutes. Do
not apply heat.
NOTE
Drying times may vary based on the environmental conditions.
Battery compartment cleaning
Under normal operation, the battery compartment should not require cleaning. If the
battery compartment does require cleaning, follow these instructions.
CAUTION
The battery compartment is not waterproof.
Do not let fluids enter the electronics through the air holes in the
battery compartment floor.
Cleaning the battery compartment in a manner other than that
specified below may cause the unit to malfunction and void the
warranty.
1. Remove the battery from the battery compartment.
2. Clean the device with a gauze pad or cloth lightly moistened with one of the
following:
Water
Soap
3. Use a cloth lightly moistened with distilled water to rinse away the cleaning
solution. Make sure moisture does not enter the electronics area below the battery
compartment floor.
4. Dry thoroughly with a lint-free cloth. Allow the battery compartment to air dry
completely prior to closing the compartment door.
Sterilization
NOTE
EtO sterilization is
not recommended
, but may be required for cables and
leadwires. Frequent sterilization will reduce the useful life of cables and
leadwires.
Sterilize with ethylene oxide gas (EtO) at a maximum temperature of 50º C (122º F).
After EtO sterilization, follow the recommendations from the sterilizer manufacturer
for required aeration.
Summary of Contents for PDM
Page 7: ...2030047 001A Patient Data Module 1 1 1 Introduction ...
Page 13: ...2030047 001A Patient Data Module 2 1 2 Equipment Overview ...
Page 21: ...2030047 001A Patient Data Module 3 1 3 Installation ...
Page 26: ...3 6 Patient Data Module 2030047 001A Installation ...
Page 27: ...2030047 001A Patient Data Module 4 1 4 Service Tool and Configuration ...
Page 38: ...4 12 Patient Data Module 2030047 001A Service Tool and Configuration ...
Page 39: ...2030047 001A Patient Data Module 5 1 5 Maintenance ...
Page 48: ...5 10 Patient Data Module 2030047 001A Maintenance ...
Page 49: ...2030047 001A Patient Data Module 6 1 6 Troubleshooting ...
Page 54: ...6 6 Patient Data Module 2030047 001A Troubleshooting ...
Page 55: ...2030047 001A Patient Data Module 7 1 7 Field Replaceable Units ...
Page 69: ...2030047 001A Patient Data Module A 1 A Appendix A Technical Specifications ...
Page 80: ...A 12 Patient Data Module 2030047 001A Appendix A Technical Specifications ...
Page 81: ...2030047 001A Patient Data Module B 1 B Appendix B Electromagnetic Compatibility ...
Page 90: ...B 10 Patient Data Module 2030047 001A Appendix B Electromagnetic Compatibility ...
Page 91: ......