Systolic pressure range
Diastolic pressure range
Mean pressure range
Cuff pressure
range
Adult: 30 to 290 mmHg
Pediatric: 30 to 240 mmHg
Neonatal: 30 to 140 mmHg
Adult: 10 to 220 mmHg
Pediatric: 10 to 200 mmHg
Neonatal: 10 to 110 mmHg
Adult: 20 to 260 mmHg
Pediatric: 20 to 215 mmHg
Neonatal: 20 to 125 mmHg
Adult: 0 to 290 mmHg
Pediatric: 0 to 250 mmHg
Neonatal: 0 to 150 mmHg
Pressure accuracy
Automatic cycle times:
Auto zero:
Tubing length:
Automatic cuff deflation:
Static: ±2% or ±3 mmHg, whichever is greater
Clinical: ±5 mmHg average error, 8 mmHg standard deviation
0 to 24 hours
Zero pressure reference prior to each cuff inflation
Variable
Cycle time exceeding 3 minutes (85 seconds neonatal),
power off, or cuff pressure exceeds 290 mmHg (
±
6 mmHg)
for adult, 250 mmHg (
±
5mmHg) for pediatric, or 150 mmHg
(
±
3mmHg) for neonatal
Cuff sizes
Disposable: Large adult, adult, small adult, pediatric, child,
and neonatal
Reusable: Adult thigh, large adult, adult, small adult, small
adult/child, child, and infant
Alarms:
User-selectable upper and lower limits for systolic, diastolic,
and mean pressures
Summary of Contents for PDM
Page 7: ...2030047 001A Patient Data Module 1 1 1 Introduction ...
Page 13: ...2030047 001A Patient Data Module 2 1 2 Equipment Overview ...
Page 21: ...2030047 001A Patient Data Module 3 1 3 Installation ...
Page 26: ...3 6 Patient Data Module 2030047 001A Installation ...
Page 27: ...2030047 001A Patient Data Module 4 1 4 Service Tool and Configuration ...
Page 38: ...4 12 Patient Data Module 2030047 001A Service Tool and Configuration ...
Page 39: ...2030047 001A Patient Data Module 5 1 5 Maintenance ...
Page 48: ...5 10 Patient Data Module 2030047 001A Maintenance ...
Page 49: ...2030047 001A Patient Data Module 6 1 6 Troubleshooting ...
Page 54: ...6 6 Patient Data Module 2030047 001A Troubleshooting ...
Page 55: ...2030047 001A Patient Data Module 7 1 7 Field Replaceable Units ...
Page 69: ...2030047 001A Patient Data Module A 1 A Appendix A Technical Specifications ...
Page 80: ...A 12 Patient Data Module 2030047 001A Appendix A Technical Specifications ...
Page 81: ...2030047 001A Patient Data Module B 1 B Appendix B Electromagnetic Compatibility ...
Page 90: ...B 10 Patient Data Module 2030047 001A Appendix B Electromagnetic Compatibility ...
Page 91: ......