Field Replaceable Units
2030047-001A
Patient Data Module
7-5
855A
4. Press the battery door closed until it seals the battery compartment.
WARNING
PHYSICAL INJURY—Make sure the battery is completely inserted
and that the battery door is securely sealed.
Falling batteries could seriously or fatally injure neonatal or other
vulnerable patients.
5. Press the
Power
button on the Patient Data Module.
6. Verify that the
Power
LED illuminates amber while the Patient Data Module
boots up, then illuminates green.
404B
Power LED
Summary of Contents for PDM
Page 7: ...2030047 001A Patient Data Module 1 1 1 Introduction ...
Page 13: ...2030047 001A Patient Data Module 2 1 2 Equipment Overview ...
Page 21: ...2030047 001A Patient Data Module 3 1 3 Installation ...
Page 26: ...3 6 Patient Data Module 2030047 001A Installation ...
Page 27: ...2030047 001A Patient Data Module 4 1 4 Service Tool and Configuration ...
Page 38: ...4 12 Patient Data Module 2030047 001A Service Tool and Configuration ...
Page 39: ...2030047 001A Patient Data Module 5 1 5 Maintenance ...
Page 48: ...5 10 Patient Data Module 2030047 001A Maintenance ...
Page 49: ...2030047 001A Patient Data Module 6 1 6 Troubleshooting ...
Page 54: ...6 6 Patient Data Module 2030047 001A Troubleshooting ...
Page 55: ...2030047 001A Patient Data Module 7 1 7 Field Replaceable Units ...
Page 69: ...2030047 001A Patient Data Module A 1 A Appendix A Technical Specifications ...
Page 80: ...A 12 Patient Data Module 2030047 001A Appendix A Technical Specifications ...
Page 81: ...2030047 001A Patient Data Module B 1 B Appendix B Electromagnetic Compatibility ...
Page 90: ...B 10 Patient Data Module 2030047 001A Appendix B Electromagnetic Compatibility ...
Page 91: ......