Appendix B: Specifications
Optima XR220amx X-Ray System
54400222-1EN
Rev 1 DRAFT 11092010
B-1
Operator Manual
© 2010 General Electric Company. All rights reserved.
Appendix B: Specifications
Power
x
Capable of 100-240V nominal, 50 /60 Hz operation
x
System battery status display
System Battery
x
Independent component power management and system battery charging algorithms to maximize
usage time
x
System Battery charge can take 90 typical exams on a single charge
x
Maximum charge time of 4 hours to go from 10-100%
X-ray Tube
X-ray Source
x
Heat Storage Capacity
–
Anode: TBD Joules (TBD heat units)
–
Tube Unit: TBD Joules (TBD heat units)
x
Maximum Heat Dissipation Rate
–
Maximum anode heat units: TBD
–
Maximum Continuous Heat Dissipation: Without Air-circulator: 170W (238 HU/s)
x
Focal Spot Sizes 1.3mm/0.6mm
x
Target Angle 12.5°
x
Anode Rotation Speed (minimal): 3200/min
x
Maximum X-ray Tube Current
–
Large Focus: 500 mA
–
Small Focus: 200 mA
x
Nominal Voltage (Radiographic): 40 to 150 kV
x
Maximum Voltage:
–
Anode to Cathode TBD kVp rectified
–
Anode or Cathode to ground TBD kVp