Section 3.7 Applying local language labels
3.6.8.3 Installing the grid
Pinch Hazard
Attach the grid holder according to
, being careful not to pinch fingers or clothing.
1.
The grid clips onto the detector. Make sure you attach it to the imaging side of the detector.
2.
With the system switched ON, verify that the grid detection operates correctly.
Each time the grid is attached, removed, or attached improperly the system alerts the user with
a double beep and displays a message on the status bar.
Section 3.7 Applying local language labels
Each Optima XR200amx/XR220amx system must be labeled with the appropriate X-ray warning
label in the local language. The system is delivered with an English label applied.
Figure 3-25 English X-ray warning label
If the system is located at a site where the primary language is not English, follow these steps.
1.
In the installation kit, locate the X-ray warning label for the local language.
2.
Clean the surface of the English label with alcohol.
3.
Allow at least two minutes for the area to dry.
4.
Apply the local language label on top of the English label.