Chapter 3 Installation
Page 47
3.5.2 Performing leakage current testing
If ground testing is required for this system, verify that the ground testing results (see
) are
< 0.1 Ohms before conducting leakage testing.
Leakage testing must be performed if the AC power plug has been replaced or disturbed.
The leakage current must not exceed 300 uA.
Before you begin, verify these steps:
1.
Complete ground testing, if required (see
3.5.1 Performing ground resistance testing on
2.
Ensure that all system covers are installed.
3.
Obtain a Dale 601/601E test meter or equivalent.
Perform leakage testing under the five specific conditions outlined in
:
Table 3-4 Leakage current test conditions
Condition
Power
Test meter
outlet
X-ray on
Section reference
1
Off
Normal
No
3.5.2.1 Condition 1: Power off, normal outlet,
2
Off
Reversed
No
3.5.2.2 Condition 2: Power off, reversed
3
On
Normal
No
3.5.2.3 Condition 3: Power on, normal outlet,
4
On
Reversed
No
3.5.2.4 Condition 4: Power on, reversed
5
On
Normal
Yes