NSTALLATION
M
ANUAL
Page 48
Section 3.5 Ground and leakage testing
Potential test equipment damage.
Always pause in the OFF position when moving the OUTLET
switch from NORMAL to REVERSED or REVERSED to NORMAL. Failure to do so can result in the
internal test meter failure and tripping of facilities circuit breakers.
3.5.2.1 Condition 1: Power off, normal outlet, X-ray off
1.
Turn off the system power, main circuit breaker ON.
On the test meter, set the Function switch to
ENCLOSURE LEAKAGE,
the L2 switch to
CLOSED
, the Outlet switch to
Normal
and the M.A.P/Lift Gnd to
OFF
(
).
Figure 3-17 Dale 601 meter