Section 2-2 - General console requirements
2-2-4
Probe Environmental Requirements
Probes should be operated, stored, or transported within the parameters outlined below.
NOTE:
See attached TEE probe user manual for the environmental requirements of 6Tc-RS probe.
NOTE:
See attached FibroScan user manual for the environmental requirements of FibroScan probe.
NOTE:
Refer to
Table 3-2 Acclimate Time on page 3-4
to determine the needed settlement time.
2-2-5
Time and Manpower Requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six
weeks before delivery, for enough time to make necessary changes.
Table 2-4
Probe Environmental Requirements
Probe
Operation
Storage
Transport
Temperature
2D probe
10
°
- 35
°
C/50
°
- 95
°
F
-40
°
- 60
°
C/-40
°
- 140
°
F
10C-D
-10
°
- 55
°
C/14
°
-131
°
F
-40
°
- 55
°
C/-40
°
- 131
°
F
4D probe
18
°
- 30
°
C/64.4
°
- 86
°
F
-10
°
- 50
°
C/14
°
- 122
°
F
Humidity
2D probe
10 - 90% non-condensing
4D probe
30 - 75% non-condensing
10 - 85% non-condensing
Pressure
2D/4D probe
700 - 1060hPa
10C-D
680 - 1060hPa
NOTICE
!! NOTICE:
WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE PRODUCT SHOULD BE KEPT
IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
CAUTION
!! CAUTION:
Have two people available to deliver and unpack the LOGIQ S8.
Attempts to move the unit considerable distances (or on an incline) by one person alone, could
result in personal injury and/or damage to the system.
Summary of Contents for LOGIQ S8
Page 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO...
Page 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x...
Page 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16...
Page 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4...
Page 417: ......