GE H
EALTHCARE
D
IRECTION
5394227, 12
LOGIQ S8/LOGIQ E8 S
ERVICE
M
ANUAL
7 - 58
Section 7-6 - Service Desktop (R4.2.5x and later)
To configure automatic data transfers
:
1.) Navigate to
ADMIN > System Admin
.
2.) On
System Admin
, configure these settings:
-
Transfer Machine Logs to GE
– Default is checked
-
Transfer Operator Information to GE
-
Transfer Images (captured without patient information) to GE
– Default is checked.
3.) Click
Save
.
4.) Navigate to
Utility (second page) > Service > Utilities > Data Transfer
.
5.) On
Data Transfer
, select
Scheduler
, and then select the days to perform the data transfer.
6.) Click
Save Settings
.
7.) To manually perform a data transfer, click
Send All
.
Summary of Contents for LOGIQ S8
Page 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO...
Page 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x...
Page 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16...
Page 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4...
Page 417: ......