Section 6-2 - Monitor Adjustment
6 - 5
6-2-1-2
Function
Figure 6-4 Function
6-2-1-2-1
Scale
1.) Press the
MENU
(middle) key of the monitor controls over 10 sec.
2.) Select the FUNCTION -> Scale by pressing the <
LEFT
or
RIGHT
> button to move the cursor and
the
MENU
(middle) key.
3.) Select FULL/5:4/NATIVE by pressing the <
LEFT
or
RIGHT
> button to decrease/increase the value
and the
MENU
(middle) key.
6-2-1-2-2
Information
1.) Press the
MENU
(middle) key of the monitor controls over 10 sec.
2.) Select the FUNCTION -> INFORMATION by pressing the <
LEFT
or
RIGHT
> button to move the
cursor and the
MENU
(middle) key.
Figure 6-5 Information