Section 9-8 - Electrical Safety Tests
9 - 23
9-8-6
Data sheet for enclosure/chassis leakage current
below shows a typical format for recording the enclosure/chassis leakage current.
Measurements should be recorded from multiple locations for each set of test conditions.
The actual location of the test probe may vary by Ultrasound system.
NOTE:
Record all data in the Electrical safety tests log.
NOTE:
Values in italics font are given as examples only.
Table 9-15 Typical data format for recording enclosure/chassis leakage
Unit under test___________________________ Date of test:_____________________
Test Conditions
Measurement/Test Point Location
System
Power Grounding/PE
Rear
Panel
Lower
Frame
Probe
Connector
Main
Handle
Off
Closed
Off
Open
On
Closed
On
Open