Section 4-3 - General procedures
4 - 11
4-3-7
Caster
4-3-7-1
Swivel/Brake Lock Caster
NOTE:
This procedure applies to RIGHT REAR Caster only.
Figure 4-9 Right Rear Caster
1.) Press ‘a’. (see:
)
2.) Rotate wheel until it locks (only one lock position in 360 deg turn).
3.) Verify Swivel Lock engages.
4.) Press ‘b’ downward. (see:
5.) Verify Brake is applied to the wheel. (i.e. Wheel does not rotate)
6.) Press ‘b’ upward.
7.) Verify Brake is released.
8.) Press ‘c’. (see:
)
9.) Verify Swivel is released.
CAUTION
!! CAUTION:
The system weighs approximately 100 kg (LOGIQ S8) / 113 kg (LOGIQ E8) with the
peripherals. Pay attention for system stability when working with the casters.
Locked
Unlocked
a
c
b