GE H
EALTHCARE
D
IRECTION
5213326-100, R
EVISION
3
LOGIQ™ A3 S
ERVICE
M
ANUAL
8-64
Section 8-7 - Power Block
Section 8-7 Power Block
8-7-1
Transformer Assembly (FRU No. 5196448)
This is a description on how to remove and replace the Transformer Assy.
8-7-1-1
Tools
•
Common Phillips screwdrivers
8-7-1-2
Needed Manpower
•
1 persons, 15 m travel
8-7-1-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-7-1-4
Removal Procedure
1.) Remove Rear cover
2.) Unscrew two screws (a,b).
3.) Remove GND cable (which ever connected near to
“a”
)
4.) Pull out the transformer tray so the connector marked J4 is accessible and disconnect Cable
(J4TRx)
CAUTION
Do not wear the ESD wrist strap when you remove a part of power supply unit. Turn OFF power
and unplug the power cord before removing a part of power supply unit. However be sure to turn
off power and wear the strap before you remove a circuit boards.
Figure 8-53 Removing Transformer Assembly
a
b
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 213 of 252