GE H
EALTHCARE
D
IRECTION
5213326-100, R
EVISION
3
LOGIQ™ A3 S
ERVICE
M
ANUAL
6-2
Section 6-3 - Monitor Adjustment
Section 6-3Monitor Adjustment
This helps to adjust the settings of the Contrast and the Brightness on the LCD Monitor suitably.
The Monitor Adjustment button is displayed as shown.
1.) Press The Toggle Button for contrast & brightness. Confirm that the Contrast (or brightness)
indicator is displayed on the monitor. If the brightness is displayed, press the toggle button again.
2.) Press the adjustment button (+/2/>) to increase contrast (or brightness)
3.) Press the Adjustment button (-/2/<) to decrease contrast (or brightness).
The amount of contrast (or brightness) is shown on a slide bar on the screen. Refer the following
table for the setting.
Record the final brightness and contrast settings and leave this information with the system. Generally
speaking, do not change the controls once they have been set, the display becomes the reference for
the hard copy device(s).
After readjusting the monitor’s Contrast and Brightness, readjust all preset and peripheral settings.
Figure 6-1 Monitor Adjustment
Table 6-2
Contrast and Brightness Recommended Setting
Room Condition
Monitor Adjustment
Contrast
Brightness
Dark room
5
4
Dim room
6
3
Bright room
7
3
room for Cardiology
8
2
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 119 of 252