GE H
EALTHCARE
D
IRECTION
5213326-100, R
EVISION
3
LOGIQ™ A3 S
ERVICE
M
ANUAL
8-54
Section 8-5 - I/O Interfaces
8-5-3
CA4-USB Cable Assy (FRU P/N # 5192651)
8-5-3-1
Tools
•
Common phillips screwdrivers
8-5-3-2
Needed Manpower
•
1person, 30 m travel
8-5-3-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-5-3-4
Removal Procedure
1. Remove Top Cover plastic
2. Remove front panel and disconnect two USB cable
3. Disconnect the top panel USB cable
4. Remove CD tray and disconnect the USB cable.
(Only if CD-ROM Option is integrated)
5. Disconnect the USB cables from the BackPlane Board