GE H
EALTHCARE
D
IRECTION
5213326-100, R
EVISION
3
LOGIQ™ A3 S
ERVICE
M
ANUAL
4-8
Section 4-4 - Functional Checks
4-4-4
Measurement and Multi Image Checks
4-4-4-1
Performing B Mode Measurements Checks
Figure 4-3 Measurement Keys
Table 4-8
Measurement Checks
Step
Task
Expected Result(s)
1.
Generate the B image.
B image is active on the screen
2.
Press Freeze
Freeze the image to measure.
3.
Press
MEASURE
on the Control Panel to
enter the Assign and Measure modality.
The measurement soft menu for the current
exam category is displayed on the Monitor.
Select the appropriate exam category and
measurement kind.
4
Trackball
the cursor
Move the cursor to the start point of the
measurement
5.
Press
Set
.
Anchor the start point of the measurement.
6.
Trackball the cursor
Move the cursor to the measurement end
point.The current distance value is displayed
in the
Measurement result table
and is
instantaneously updated when moving the
cursor
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 95 of 252