GE H
EALTHCARE
D
IRECTION
5213326-100, R
EVISION
3
LOGIQ™ A3 S
ERVICE
M
ANUAL
Chapter 10
10-9
10-5-4
Cleaning
10-5-4-1
General Cleaning
10-5-4-2
Air Filter Cleaning
NOTE:
For your convenience or of the air filter is too dirty, replacement filters are available. Refer to
Chapter 9 for the air filter replacement part number.
Table 10-10 General Cleaning
Step
Item
Description
1
Console
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system. Be careful
not to get the cloth too wet so that moisture does not enter the console.
2
Probe Holder Clean probe holders (they may need to be soaked to remove excess gel).
3
Monitor
LCD monitor, use a non-ammonia (just isopropyl and water) lens cleaner. These are available at most
computer outlet stores. DO NOT use Windex, Screen-Clean, etc., because these contain ammonia, which
will remove the anti-glare coating on the monitor.
4
Keyboard
(Rubber
Keypad)
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system. Be careful
not to get the cloth too wet so that moisture does not enter the console.
Table 10-11 Air Filter Cleaning - frequency varies with your environment
Step
Item
Description
1
Remove Filter
Cover
Refer to Chapter 8 for air filter location and removal instructions.
2
Clean Filter
The filters can be cleaned in sprinkling water, or they can be dusted with a vacuum cleaner. If the filter
is metal wash and/or vacuum. If the filter is fiber or plastic vacuum or replace.
3
Install Filter
Install the clean filter.