GE H
EALTHCARE
D
IRECTION
5213326-100, R
EVISION
3
LOGIQ™ A3 S
ERVICE
M
ANUAL
Chapter 7
7-23
Section 7-6 Service Diagnostics
Figure 7-8 Service diagnostics Login Screen
4. Enter User Name:
Service
Password:
Sample
(Confidential) and press OK.
5. Once you type the User name and password, service diagnostics main screen (Refer Figure 7-4) will
appear.
Main Screen Consists of the following buttons:
1.)
Full Tests:
To do the tests on MST, TRX, and CON boards.
2.)
KeyBoard Test:
To perform the Keyboard Test
3.)
LCD Test:
To perform the Display Tests
4.)
Error Logs:
To view the Errors occurred in the system
5.)
Event Logs:
To view the Event Logs .
6.)
Copy to USB:
To copy the Log files to USB
7.)
Copy to CD:
To copy the Log files to CD
8.)
Return:
Return to the main Utility screen.
NOTE:
For all kind of Tests
If Indicator is Red ------Test is failed and you may see Error Logs .
If indicator is Green ------ Test is passed and all the components are Ok.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5213326-100TPH_r3.pdf Page 142 of 252