LOGIQ 7/LOGIQ 7 Pro Quick Guide
Direction 5307393-100 Rev. 1
49
Safety
Peripheral Update for EC countries
General precautions for installing an alternate off-board, remote device or a network would include:
1.
The added device(s) must have appropriate safety standard conformance and CE Marking.
2.
The added device(s) must be used for their intended purpose having a compatible interface.
3.
Signal or mains isolation devices and additional protective earth may be needed to assure compliance with IEC/EN 60601-1-1.
Declaration of Emissions
This system is suitable for use in the following environment. The user must assure that it is used only in the electromagnetic environment as specified.
CAUTION
The connection of equipment or transmission networks other than as specified in the user instructions can result in an electric shock
hazard or equipment malfunction. Substitute or alternate equipment and connections requires verification of compatibility and conformity to
IEC/EN 60601-1-1 by the installer. Equipment modifications and possible resulting malfunctions and electromagnetic interference are the
responsibility of the owner.
Table 1-6: Declaration of Emissions
Guidance and manufacturer’s declaration - electromagnetic emissions
The system is intended for use in the electromagnetic environment specified below. The user of the system should assure that it is used in such an environment.
Emission Type
Compliance
Electromagnetic Environment
RF Emissions
CISPR 11
Group 1
This system uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF Emissions
CISPR 11
Class A
This system is suitable for use in all establishments, other than domestic establishments and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes, provided the
following
warning is heeded:
Warning: This system is intended for use by healthcare professionals only. This system may cause radio interference or may disrupt the operation of nearby
equipment. It may be necessary to take mitigation measures, such as re-orienting or relocating the
system
or shielding the location.
Harmonic
Emissions
IEC 61000-3-2
Class A
Voltage
Fluctuations/Flicker
Emissions
IEC 61000-3-3
Complies