LOGIQ 7/LOGIQ 7 Pro Quick Guide
Direction 5307393-100 Rev. 1
24
Connectivity
Connectivity on the LOGIQ 7/LOGIQ 7 Pro is based on
the Dataflow concept.
Login as Administrator. Select
Utility
. Select
Connectivity
. Configure the menus from left to
right, starting with TCP/IP first. When finished
making connectivity changes to the utility menus,
restart the system.
TCP/IP
Type in the Computer’s Name. Identify the Ultrasound
system to the rest of the network by filling in its IP
Address, Subnet Mask, and Gateway (if applicable).
Press
Save
.
Device
1.
Press
Add
.
2.
Type the name of the device and its IP address.
3.
Press
Save
.
Services
To add the services,
1.
Select the destination device.
2.
Select the service to add and press
Add
.
3.
Type the name in the Name Field.
4.
Specify Properties in the Properties boxes (located in
the upper right-hand side and lower left-hand side).
5.
Press
Save
.
There are two service types that pertain to printers:
•
Standard Printer is used for digital peripherals.
•
Video Capture Device is used for devices that are
triggered by a contact closure, typically analog
devices.