LOGIQ 7/LOGIQ 7 Pro Quick Guide
Direction 5307393-100 Rev. 1
47
Safety
Notice upon Installation of Product
Separation distance and effect from fixed radio communications equipment: field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast transmitter cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the ultrasound system is used exceeds the applicable RF compliance level as stated in the immunity declaration, the ultrasound system should be
observed to verify normal operation. If abnormal operation is observed, additional measures may be necessary, such as re-orienting or relocating the ultrasound
system or using an RF shielded examination room may be necessary.
1.
Use either power supply cords provided by GE Medical Systems or ones designated by GE Medical Systems. Products equipped with a power source plug
should be plugged into the fixed power socket which has the protective grounding conductor. Never use any adaptor or converter to connect with a power
source plug (i.e. three-prong-to-two-prong converter).
2.
Locate the equipment as far away as possible from other electronic equipment.
3.
Be sure to use only the cables provided by or designated by GE Medical Systems. Connect these cables following the installation procedures (i.e. wire power
cables separately from signal cables).
4.
Lay out the main equipment and other peripherals following the installation procedures described in the Option Installation manuals.
General Notice
1.
Designation of Peripheral Equipment Connectable to This Product.
The equipment indicated on
Chapter 15
of the
Basic User Manual
can be hooked up to the product without compromising its EMC performance.
Avoid using equipment not designated in the list. Failure to comply with this instruction may result in poor EMC performance of the product.
2.
Notice against User Modification
The user should never modify this product. User modifications may cause degradation in EMC performance.
Modification of the product includes changes in:
a.
Cables (length, material, wiring, etc.)
b.
System installation/layout
c.
System configuration/components
d.
Securing system parts (cover open/close, cover screwing)
3.
Operate the system with all covers closed. If a cover is opened for some reason, be sure to shut it before starting/resuming operation.
4.
Operating the system with any cover open may affect EMC performance.