MA-041
–
DGA 900 Plus - Installation & Commissioning Manual - Rev 1.1
4-Mar-22
Page 43 of 99
8.
Connect the cable to the terminals
of the HFCT.
9.
The central white wire must be
connected to the
core
side.
10.
The braid shield must be
connected to the
shield
side.
11.
Tighten the set screws using a hex
key wrench (1.5 mm).
12.
Screw the fitting body into the
HFCT with the seal ring. Hold the
wire to prevent it from twisting.
Recommended torque is 3.75 N m
(33 lb in).
13.
Screw the dome nut tightly into
place. Recommended torque is
2.50 N m (22 lb in).
14.
The fitting and the nut must be
sufficiently tight to prevent water
ingress.