MA-041
–
DGA 900 Plus - Installation & Commissioning Manual - Rev 1.1
4-Mar-22
Page 62 of 99
Figure 9-40:
Configure > Save last measured data > as “load” calibration
Select
File > Save As
and type the filename
calibLoad
as shown in Figure 9-41 to
save the data in the folder created in Section 9.8.2.1 for this Bushing Adaptor.
Figure 9-41:
Save As “calibLoad”
9.8.2.2.4
Adaptor-only Test
Connect the clips to the Bushing Adaptor using the short cables in the adaptor’s
connector as shown in Figure 9-42.