Hydran* M2-X Instruction Manual
Page 30
MA-029
Rev. 1.0,
30-Jul-18
5.
Expansion port
: Not used. This is used in Hydran S2 only.
Figure 2-6: Left Side of the Electronic Card Cage
2.3.4
Inside
The electronic card cage contains the circuit boards (CPU, sensor card, etc.) and
the battery.
2.3.4.1
Battery
The Hydran M2-X
’s battery is used to keep the real
-time clock functioning and to
retain the parameter values and data of the history files when the ac power is not
available. For the specifications, see Section 9.3; should you need to replace the
battery, see Appendix G.
2.3.4.2
Internal Real Time Clock
An internal real-time clock (RTC) provides a calendar and a permanent clock,
unless the battery is disconnected. The clock uses a crystal oscillator as its time
base; its 0.01 % accuracy (four minutes per month) is independent of the ac power
supply frequency. The calendar accounts automatically for leap years (366-day
year).
To set the Hydran M2-X
’s date and time,
see Section 3.4.3.
The Hydran M2-X
’s internal clock manages the following functions:
1. Temporary
Plug
2. AC Power Supply
Terminal Block
3. AC Power Supply
Fuse
4. Alarm Relays
Terminal Block
5. Expansion Port