Hydran* M2-X Instruction Manual
Page 126
MA-029
Rev. 1.0,
30-Jul-18
-
Batt. L
;
SysOK Relay;FaultTrig.
submenu; see Section 3.2.3.
-
Batt.Alr L
(alarm set point);
Service;VoltPile
submenu; see Section 3.3.2.4.
-
BattAlr Delay
;
Service;VoltPile
submenu; see Section 3.3.2.4.
•
Alarm cause(s)
: Battery voltage below the
Alarm Low SP
alarm set point
.
Replace the battery within the next nine months. For details, see Appendix G.
8.1.3
B.P. Temp Hi Alarm
•
Alarm type
: Due to system fault conditions.
•
Non-abbreviated description
: High base plate temperature.
•
Alarm message on the Perception software
: Base Plate Temp High.
•
Priority
: High.
•
Alarm cause(s)
: Temperature of the base plate above the
Alarm Hi SP
alarm
set point.
8.1.4
B.P. Temp Low Alarm
•
Alarm type
: Due to system fault conditions
.
•
Non-abbreviated description
: Low base plate temperature
.
•
Alarm message on the Perception software
: Base Plate Temp Low
.
•
Priority
: High
.
•
Alarm cause(s)
: Temperature of the base plate below the
Alarm Low SP
alarm
set point
.
8.1.5
B.P. Tmp C. Open Alarm
•
Alarm type
: Due to system fault conditions
.
•
Non-abbreviated description
: Low-Low base plate temperature
.
•
Alarm message on the Perception software
: Base Plate Temp Cable Open
.
•
Priority
: High
.
•
Alarm cause(s)
: Temperature of the base plate below the
Alarm Lo-Lo SP
alarm set
point.
8.1.6
B.P. Tmp C. Short Alarm
•
Alarm type
: Due to system fault conditions
.
•
Non-abbreviated description
: High-High base plate temperature
.
•
Alarm message on the Perception software
: Base Plate Temp Cable Short
.
•
Priority
: High
.