Chapter 8 - Replacement Procedures
8-75
8-4-9-4
Peripheral Power Cable Installation Procedure
Correct routing of the Peripheral AC Outlet cable is shown in
NOTE:
Before starting the routing procedure adhere two plastic cable holders to the metal linear guide bracket
as shown in
(middle picture).
1.) Insert the replacement Peripheral Power cable and power socket in position in the SafeLock’s
Upper Cover (front) (see item 4 in
2.) With the AC power socket plugged into the cart (
, left picture), connect the ground cable
to the body of the cart.
3.) Route the Peripheral AC Outlet cable as shown in the middle picture of
.
4.) Secure the cable with the cable clamp in the direction shown in
5.) Route the cable and the ground cable over the plastic cable holders (
middle picture)
and secure the cables with the tie-wraps.
Figure 8-86 Routing of Peripheral AC Outlet Cable from Left to Right
Figure 8-87 Correct Installation of Cable Clamp
Summary of Contents for FL000090
Page 2: ......
Page 30: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL xxviii ...