Section 6-2 - LCD Display Adjustments and Calibration
6-2-5-1
Adjusting the Blue Tint
NOTE:
The default blue tint test pattern is set at two bars - see
1.) Press
Alt Q
to access the Display Calibration Screen.
2.) To adjust the blue tint: hold down <ALT>, then use the < and > keys, as required.
Repeatedly pressing < will decrease the blue tint; repeatedly pressing > will increase the blue tint.
Note:
To assist in making blue tint adjustments, an adjustment bar is displayed below the screen
- see
.
3.) Press
Alt Q
to exit to the Display Calibration Screen.
Figure 6-7 LCD Display Calibration Screen showing Test Pattern
Figure 6-8 Blue Tint Adjustment Bar