GE H
EALTHCARE
D
IRECTION
FQ091013, R
EVISION
1
V
IVID
I
N
AND
V
IVID
Q
N S
ERVICE
M
ANUAL
10-14
Section 10-5 - System Maintenance
10-5-4-2
Disinfection
The disinfection procedures described in this section should be routinely performed by medical
personnel, as shown in
. The information is provided in this manual for reference purposes.
NOTICE
IMPORTANT - To avoid contamination when the system is in use during operations, always
cover the keyboard and the monitor with the OR covers provided for this purpose.
Table 10-9
Routine Disinfection Procedures
What?
When?
How?
Materials?
Who?
Keyboard and Trackball
Weekly.
In event of
contamination,
clean immediately
with a 70:30
mixture of isopropyl
alcohol and water.
Switch the system off!
Disinfectant with a
70:30 mixture of
isopropyl alcohol and
water.
Never allow liquid to
flow into the keyboard!
Use a 70:30
mixture of
isopropyl
alcohol and
water.
Staff users/doctors
Monitor display-surface
Weekly.
In event of
contamination,
clean immediately
with a 70:30
mixture of isopropyl
alcohol and water.
Clean with dry soft,
lint-free cloth.
When contaminated
with blood or other
stains secretions, use a
70:30 mixture of
isopropyl alcohol and
water.
Soft,
lint-free cloth.
Disinfect with a
70:30 mixture of
isopropyl
alcohol and
water.
Staff users/doctors
Devices' external surfaces
(incl. monitor housing, handle
and SafeLock Cart wheels)
Weekly.
In event of
contamination,
clean immediately
with a 70:30
mixture of isopropyl
alcohol and water.
Clean the washable
outer surfaces with a
soft, non-abrasive cloth
including the SafeLock
Cart wheels. Do not use
abrasive powder or
strong
chemicals!
Use a soft,
non-abrasive
folded cloth or
sponge with a
mild, general
purpose,
non-abrasive
soap and water
solution.
Trained cleaning staff
Summary of Contents for FL000090
Page 2: ......
Page 30: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL xxviii ...
Page 739: ......