User Manual
Druck DPI612
Page 94 of 105
[EN] English – 109M4017 Revision -
6
ACCESSORY INSTRUCTIONS
Dirt Moisture Trap 20 bar (P/N IO620-IDT621)
6.1
6.1.1
Specification:
Maximum working pressure: 20bar / 300psi pneumatic.
Orientation: Use only in the vertical position.
6.1.2
Pressure connections:
Lower: To DPI612 Test Port.
Upper: AMC quick-connect ( for use with all DPI612 adaptor sets )
Materials of Construction: 316L and 303 Stainless steels, Acrylic,
Nitrile seals, P.T.F.E.
6.1.3
Operation:
Important Note:
This device is designed for use on the DPI 612 pFlex 20 bar
pneumatic pressure calibrator.
1.
Fit the Dirt Moisture Trap to the Test Station of the DPI 612
pFlex.
2.
If moisture or particles are seen within the transparent
chamber during use, dismantle and clean the trap. ( see
below).