[EN] English – 109M4017 Revision -
4.
Use
to stop the Leak Test.
2.5.2
Set the Pressure Module to Zero
Use this option to write a new zero pressure value to the pressure
module in use. The sensor adjustment is permitted if it obeys the
condition that follows:
•
Adjustment ≤10% FS positive pressure value (for the Sensor).
Note
: To make a temporary adjustment for zero, use the Tare
function.
2.5.3
Error Indications
Under range: The display shows this symbol for this condition:
<<<<<
Reading < 110% Negative Full Scale (Pressure)
Reading < 102% Negative Full Scale (Electrical)
Over range: The display shows this symbol for this condition:
>>>>>
Reading > 110% Positive Full Scale (Pressure)
Reading > 102% Positive Full Scale (Electrical)
If the display shows <<<< (under range) or >>>> (over range):
•
Make sure that the range is correct.
•
Make sure that all the related equipment and connections are
serviceable.
MEASURE PRESSURE: IDOS Option
2.6
Optional item, An IDOS Universal Pressure Module (UPM) uses
Intelligent Digital Output Sensor (IDOS) technology to measure the
applied pressure and supply the data to an IDOS instrument. Before
using an IDOS module, (Ref: User Manual: K0378, Druck IDOS UPM).
Note:
To attach an IDOS module to the Druck DPI612 use an IO620-
IDOS-USB adaptor.