User Manual
Druck DPI612
Page 23 of 105
[EN] English – 109M4017 Revision -
DPI 612 pFlexPro
(100 bar) Pump
When you have set the operation to pressure or vacuum (
See
Section 1.9.4
), seal the system (
See Section 1.9.3
) and use the
pump to set the necessary pressure or vacuum.
You can then make the last adjustments with the volume
adjuster (
See Section 1.9.6
).
Figure 1-8
100 bar Pump
DPI 612 hFlexPro (1000 bar) Pump
When you have set the operation to pressure or vacuum
(
See Section 1.9.4
), seal the system (
See Section 1.9.3
) and
use the pump to set the necessary pressure or vacuum.
You can then make the last adjustments with the volume
adjuster (
See Section 1.9.6
Figure 1-9
1000 bar
Pump
1.10.6
Volume Adjuster
DPI 612 pFlex (20 bar) Volume Adjuster
This control increases or decreases the pressure/vacuum.
Before you seal the system (
See Section 1.9.3
), turn this
control to the necessary position:
•
For equal adjustment, turn it to the middle of
its range.
•
For maximum adjustment, turn it fully
clockwise or counter clockwise.
When you have set the necessary pressure or vacuum
with the pump (
See Section 1.9.5
), use the volume adjuster
to make the fine adjustments.
Figure 1-10
20 bar
Volume
Adjuster